Management of radiation dose reduction in computed tomography: an experience at burapha university hospital, thailand

Limchareon S

Division of Radiology and Nuclear Medicine, Faculty of Medicine, Burapha University, Thailand

E-mail : aa

Kaowises K

Division of Radiology and Nuclear Medicine, Faculty of Medicine, Burapha University, Thailand

Saensawas W

Division of Radiology and Nuclear Medicine, Faculty of Medicine, Burapha University, Thailand

DOI: 10.15761/RDI.1000123

Article
Article Info
Author Info
Figures & Data

Abstract

Objective: The purpose of this study was to quantify the dose reduction for each computed tomography (CT) reduction doseprotocol.

Subject and Method: One hundred and fifty repeat CT examinations with an implementation of a CT dose reduction technique per examination were included. Percent of dose reductions compared with their prior CT examinations with standard CT protocols in terms of dose length productwere calculated.

Result: There were 95 examinations using scanning parameter optimization technique, 47 examinations using reduction of multiphase scanning and 8 examinations using limited scan range. Percent of dose reductions for scanning parameter optimization, reduction of multiphase scanning and limitation of scan range were43.3%, 52.5%, and 39.1% respectively. Among the scanning parameter optimization, percent of dose reductions by lowering tube voltage, and lowering tube current were 47.9%, and 38.6%, respectively.

Conclusion: The presented dose lowering protocols are simple and available in most modern CT scanners. Reduction of multiphase scanning shows highest dose reduction. All of these techniques do not depend on CT scanners but requiring special CT protocol tailored for the individual patient for optimal diagnostic quality in the optimizationof patient dose in CT.

Keywords

computed tomography, dose reduction, radiation dose

Abbreviations

ACR - American college of radiology;CT- computed tomography; DLP - dose length product; kVp-kilovotage peak; mA – milliampere;MDCT - multi-detector computerized tomography;IR - iterative reconstruction

Introduction

 Since the era of multi-detector computerized tomography (MDCT), the rate of computed tomography (CT) examinations in our institution has been increasing rapidly, similar to another part of the world[1].Although MDCT has a lot of benefits, the radiation dose from MDCT is higher than other imaging modalities [2]. Thus, the risk of carcinogenesis from radiation is to be concerned[2].A retrospective study in a large group of patients published in 2012 showed significantassociations between radiation doses from CT scans and subsequent incidence of leukemia and brain tumors in children aged younger than 15 years [3].But awareness of this possible risk is less concerned among physician as demonstrated by a survey inemergency department (ED)physicians that, only 9% of them believed that radiation increases cancer risk [4].Even radiologists in that study, less than 50% of thembelieved that radiation increases cancer risk and most of them were unable to correctly estimate the dose for one CT scan compared with a chest x-ray [4]. Similarly, most ED physiciansand medical students were unable to correctly estimate the dose for one CT scan compared with a chest x-ray [4-5]. In addition, almost 80% of medical students and interns who are future physicians have never heard of the American College of Radiology (ACR) appropriateness criteria[5-6].

Many efforts have tried to minimizing radiation dose from CT either by vendors or users. All vendors have available dose reduction technology and some of the technologies are being in research. Whereas CT users can minimize radiation dose by scanning parameter optimization, reduction of inappropriate multiphase scanning, limiting the scan range in the Z-axis, and applying innovative technologies such as iterative reconstruction (IR) algorithms that are available in some new CT scanners.

Our institution has installed a 64-slice MDCT (Toshiba Aquilian 64; Otawara, Japan) since April2013.Initially, we used standard protocols for all of CT examinations. After a period of time, we have noticed that there were a number of follow-up cases in our institution, and rapidly increasing as shown in figure 1. Then CT protocols for reducing radiation dose in such cases have been implemented.

Figure 1. Number of follow-up cases at Burapha University hospital.

The aim of this study was to measure and express the amount of dose reduction in different CT techniques.We compared low-dose CT protocol with standard-dose CT protocol in the same patient. We do not aim to compare the statistically significantdifference between these techniques, because all of these techniques should be used, even in the single examination.

Material and method

This study was retrospectively designed in a 160-bed university hospital. The university institutional review board approved this study, No.17/2560. Informed consent was waived due to retrospective nature of the study.

We searched the CT database for all patients who underwent repeat CT examinations and received dose-lowering protocols in the follow-up examinationssince the CT installation on April 2013 toDecember2016.A total of 150 CT examinations were identified. The CT procedures were performed for the clinical indications of follow-up in patients as illustrated in table 1.Radiation dose estimates were determined using the dose-length product (DLP) in mGy-cm. The DLP is a measurement of the radiation delivered by a CT scanner for each examination multiplied by scan length[7]. The DLP which is displayed on the scanner console was used for comparing radiation dose per examination ina single patient.Each repeat examination was obtained at almost the identical anatomic location to the prior examination. The SPSS version 22 was used for statistical analysis. All results are reported as frequency, percent, mean and standard deviation as appropriate. Paired t-test was used to compare the radiation doses between the first and the second examinations.

Table 1. Indications for follow up.

Dose lowering techniques

Indications

No. of patients

1.Optimized technical scanner settings

 

    1.1 Decreased tube current

Intracranial hematoma/infarct/mass

72

Urinary tract stone/cyst

8

Abdominal cancer

1

    1.2 Decreased tube potential

Urinary tract stone

12

Post-abdominal aortic aneurysm stent

1

Pneumomediastinum

1

2.Limited image creep

Abdominal cancer

3

Renal nodule/cyst

2

Intra-abdominal collections

2

C-spine fracture

1

3.Limited phase creep

 

    3.1 omitted 1 phase

Abdominal cancer

22

Liver nodule/cyst

2

Hepatobiliary infection

2

Intra-abdominal collections

1

Intra-abdominal foreign body

1

Kidney injury

1

    3.2 omitted 2 phases

Abdominal cancer

5

Liver nodule/cyst

3

Kidney injury

1

Post-operative pain

1

Intra-abdominal collections

1

Adrenal gland nodule

1

Pulmonary tuberculosis

1

    3.3 omitted 3 phases

Abdominal cancer

2

Adrenal gland nodule

2

Intra-abdominal infection

1

150

CT dose lowering protocols

The initial CT examinations were performed using standard CT technical scanner settings (helical mode and fixed tube potential at 120 kilovoltage peak [kVp]). The head and neck CT examinations were performed using fixed tube current at 300milliamperes (mA), 0.75second (s) rotation time, pitch 0.8 and 2 mm reconstruction interval. The chest, abdomen and KUB (Kidney-Ureter-Bladder) CT were done, using tube current modulation[8]275-500 mA, 0.5s rotation time, pitch 0.8 and 2 mm reconstruction interval.The dose lowering techniques included; 1) optimized technical scanner settings, 2) limited image creep, and 3) limited phase creep. Of the optimized technical scanner settings, we subdivided into loweringtube currentand loweringtube potential. Among lowering tube current, we reduced the tube current to 225 mA, in cases of fixed tube current whilst in cases of using tube current modulation, we limited the maximum range of tube current to 400 mA. In cases of lowering tube potential, we reduced to 100kVp and fixed tube current at 250 mA.Of the limited image creep, we limited image creep to only the region of interest, either on non-contrast phase or on delayed phase. Similarly, in the limited phase creep, we omitted one, two or three phases, usually anon-contrast phase, and/or late-arterial phase, and/or delayed phase. In a single patient, we used only one technique per one examination whereas the other parameters were held the same as the prior examination.

Result

The study included 150patients, mean age 53.2years (range, 13 years to 88years). Of whom, 62 were female (41.3%) and 88 were male (58.7%).Thenumber of patients, mean difference and percentage of dose reduction values with respect to dose-lowering techniques are presented in table 2. The highest dose reduction technique was for the limited phase creep (average 52.5%), while the lowest dose reduction technique was for thedecreased tube current (38.6%).

Table 2. Mean difference and percentage of dose reduction.

n = number of examinations, SD = Standard deviation, * = Statistical significance

Dose lowering technique

(n)

Mean difference (SD)

95% Confidence interval

Dose reduction (%)

p

1.Optimized technical scanner settings

 

   1.1 decreased tube current (81)

526.6 (470.1)

422.7 – 630.6

38.6

<0.001*

   1.2 decreased tube potential (14)

643.8 (404.8)

409.4 - 876.9

47.9

<0.001*

2.Limited image creep (8)

1152.9 (413.1)

807.5 – 1498.3

39.1

<0.001*

3.Limited phase creep

 

 

 

   3.1 omitted 1 phase (29)

1295.9 (721.6)

1021.4 – 1570.4

36.7

<0.001*

   3.2 omitted 2 phases (13)

2767.2 (1437.0)

1898.8 – 3635.5

58.7

<0.001*

   3.3 omitted 3 phases (5)

3103.9 (1779.2)

894.7 – 5313.0

62.1

0.018*

Discussion

Although current evidence is sufficient that, radiation increases the risk of cancer [2-3], benefits from CT in timely and correct diagnosis have increased the usefulness of CT[1]. CT radiation doses depend on patient size, scan coverage, imaging parameters, andCT scanner technology. The current study demonstrates that the limited scan phase had the highest dose reduction (52.5%). The un-indicated multiphase scansare usually overlooked by staff, according to the study by Guite et al [9]. They studied abdominal and pelvic CT examinations and reported that 52.8% of phases were not supported by ACR appropriateness criteria and 35.8% of phases were un-indicated. Limited scan phase technique is very simple, no image quality penalty and most importantly,this technique does not depend on CT scanners.

Limited image creep is one of the simplest techniques for minimizing radiation, especially in follow-up patients. The patients are previously diagnosed and the region of interest that is required is specifically scanned. Individualize CT scans according to the region of interest by theradiologist is mandated.

For the optimized technical scanner settings included; lowering tube current, loweringtube potential, and increased pitch, all ofthese techniques result in image deteriorationby increasing noise [8].Therefore, image quality and exposure dose reduction are a tradeoff.Punwaniet al.[10]studied in phantoms for pulmonary nodule detection. They demonstrated that the sensitivity and specificity for pulmonary nodule detection in phantoms decreased when the noise increased. As shown by the study in pulmonary embolism that, the simulated reduction in CT doses had a significant adverse effect on the readers’ diagnostic confidence and image quality [11]. However, in clinical practices, authors achieved an excellent diagnostic image quality with alow dose and ultra-low-dose techniques in the diagnosis of infectious lung disease[12] as well as an excellent sensitivity (97%) and specificity (96%) for diagnosing renal colic [13]. A recent article with a large number of patients and various clinical settings, they showedsignificantly lower CT radiation doses to 53%-73% while maintained acceptable diagnostic image quality [14].Though in that study, the groups of patients between before and after dose-reduction implementation were not the same and two reduction techniques were used at the same time. Previous literature has proposed a reconstruction algorithm to provide diagnostic quality images in minimizing radiation dose settings [15]. They demonstrated the benefits of resampling the scan data by resampling 5 mm thick slice acquisitions to 0.6 mm slices and then producing 2 mm coronal MPRs, that improved the spatial resolution but with a moderate noise penalty. Recently, objective data have shown that the use IR significantly reduced noise, compared to no IR with the other identical parameters (p<0.017) [16].However, IR is innovative technology and is not available in our CT scanner.

The tube current (mA) modulation is the dose reduction technique which is usually available in the modern MDCT scanners and already set in the standard CT protocols[8]. The scan mA value is adjusted, along with the length of thepatient, for each tube rotation. The mA modulation technique has been reported to reduce 40% of radiation dose per examination[8]. In the current study, we limited the maximum range of the variable mA. As a result, we can further reduce the radiation dose to38.6%. Decreasing tube current has a direct linear effect on the decreasing radiation does whereas the radiation dose does not decrease proportionally to the tube potential [17]. We decreased tube potential from 120 kVp to 100 kVp and the result showed 47.9% of dose reduction.However, decreased tube potential is not recommended in thelarge patient because beam attenuation increases exponentially with increasing patient thickness [18]. Recently, there has been automatic tube potential selection technique in some vendors[17].

Radiologist’s bias is the major limitation of our study. The dose-lowering CT protocols were initiated by an author (L.S) who interpreted the CT images. In addition, these CT examinations were follow-up examinations and the radiologist were aware of the previous results.This may not be generalized to the other clinical indications. We did not try to compare the image quality or accuracy which is an important effect on the diagnostic outcome. Further assessment of the accuracyof each technique is required. Some dose-lowering protocols have considerable image noise [19], therefore the usability may be limited to some diagnostic questions. The DLP that we used to represent the radiation dose does not the true patient effective dose. Lastly,a small number of patients often limited the study results.

In conclusion, we present ready-to-use types of radiation dose-lowering techniques. These techniques are simple and likely available in most CT scanners. The limited phase creep technique reduces highest radiation dose whereas the reduced tube current is the least. We would like to emphasize on radiologists’ awareness and promote all of these techniques to be used in daily practice especially in patients who receive repeat imaging.Nevertheless, these techniques require a special protocol tailored for the individual patient and thus need focus and change in the radiologists’ practices.

Acknowledgement

The authors are grateful to Faculty of Medicine, Burapha University for all supports.

Conflict of interest

No conflict of interest to disclose.

References

  1. Fletcher JG, Kofler JM, Coburn JA, Bruining DH, McCollough CH (2013) Perspective on radiation risk in CT imaging. Abdom Imaging 38: 22-31.[Crossref] 
  2. Sodickson A, Baeyens PF, Andriole KP, Prevedello LM, Nawfel RD, et al. (2009) Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology 251: 175-184.[Crossref]
  3. Pearce MS, Salotti JA, Little MP, McHugh K, Lee C, et al. (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet380: 499-505.[Crossref]
  4. Lee CI, Haims AH, Monico EP, Brink JA, Forman HP (2004) Diagnostic CT scans: assessment of patient, physician, and radiologist awareness of radiation dose and possible risks. Radiology 231: 393-398.[Crossref]
  5. Prezzia C, Vorona G, Greenspan R (2013) Fourth-year medical student opinions and basic knowledge regarding the field of radiology. AcadRadiol 20: 272-283.[Crossref]
  6. Saha A, Roland RA, Hartman MS, Daffner RH (2103) Radiology medical student education: an outcome-based survey of PGY-1 residents. AcadRadiol 20: 284-289.[Crossref]
  7. Goldman LW (2007) Principles of CT: radiation dose and image quality. J Nucl Med Technol 35: 213-225.[Crossref] 
  8. Mayo-Smith WW, Hara AK, Mahesh M, Sahani DV, Pavlicek W (2014) How I do it: managing radiation dose in CT. Radiology 273: 657-672.[Crossref]
  9. Guite KM, Hinshaw JL, Ranallo FN, Lindstrom MJ, Lee Jr FT (2011) Ionizing radiation in abdominal CT: unindicated multiphase scans are an important source of medically unnecessary exposure. J Am Coll Radiol8: 756-761.[Crossref]
  10. Punwani S, Zhang J, Davies W, Greenhalgh R, Humphries P (2008) Paediatric CT: the effects of increasing image noise on pulmonary nodule detection. PediatrRadiol 38: 192-201.[Crossref] 
  11. MacKenzie JD, Nazario-Larrieu J, Cai T, Ledbetter MS, Duran-Mendicuti MA, et al. (2007) Reduced-dose CT: effect on reader evaluation in detection of pulmonary embolism. AJR Am RoentgenolRadiol189: 1371-1379.[Crossref]
  12. Rizzi EB, Schinina V, Gentile FP, Bibbolino C(2007) Reduced computed tomography radiation dose in HIV-related pneumonia: effect on diagnostic image quality. Clin Imaging31: 178-184.[Crossref]
  13. Poletti PA, Platon A, Rutschmann OT, Schmidlin FR, Iselin CE, et al. (2007) Low-dose versus standard-dose CT protocol in patients with clinically suspected renal colic. AJR Am RoentgenolRadiol 188: 927-933.[Crossref]
  14. Sulagaesuan C, Saksobhavivat N, Asavaphatiboon S, Kaewlai R (2016) Reducing emergency CT radiation doses with simple techniques: a quality initiative project. J Med Imaging Radiat Oncol60: 23-34.[Crossref]
  15. Ford NL, Protik A, Babyn P, Thomas K (2014) Does resampled image data offer quantitative image quality benefit for pediatric CT? J Biomed Sci Eng7: 343-350.
  16. Higashigaito K, Becker AS, Sprengel K, Simmen HP, Wanner G, et al. (2016) Automatic radiation dose monitoring for CT of trauma patients with different protocols: feasibility and accuracy. ClinRadiol 71: 905-911.[Crossref]
  17. Lira D, Padole A, Kalra MK, Singh S (2015) Tube potential and CT radiation dose optimization. AJR Am RoentgenolRadiol204: W4-W10.[Crossref]
  18. Moser JB, Sheard SL, Edyvean S, Vlahos I (2017) Radiation dose-reduction strategies in thoracic CT. ClinRadiol 72: 407-420.[Crossref] 
  19. Hojreh A, Czerny C, Kainberger F (2005) Dose classification scheme for computed tomography of the paranasal sinuses. Eur J Radiol56: 31-37.[Crossref]

Editorial Information

Editor-in-Chief

Yi-Hwa Liu
Yale University

Article Type

Research Article

Publication history

Received date: January 02, 2018
Accepted date: January 19, 2018
Published date: January 23, 2018

Copyright

©2018 Limchareon S. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use,distribution, and reproduction in any medium, provided the original author and source are credited.

Citation

Limchareon S, Kaowises K, Saensawas W (2018) Management of radiation dose reduction in computed tomography: an experience at burapha university hospital, thailand.Radiol Diagn Imaging 2: DOI: 10.15761/RDI.1000123

Corresponding author

Limchareon S

Division of Radiology and Nuclear Medicine, Faculty of Medicine, Burapha University,Thailand; Tel: +6638386554; Fax: +6638386557

Table 1. Indications for follow up.

Dose lowering techniques

Indications

No. of patients

1.Optimized technical scanner settings

 

    1.1 Decreased tube current

Intracranial hematoma/infarct/mass

72

Urinary tract stone/cyst

8

Abdominal cancer

1

    1.2 Decreased tube potential

Urinary tract stone

12

Post-abdominal aortic aneurysm stent

1

Pneumomediastinum

1

2.Limited image creep

Abdominal cancer

3

Renal nodule/cyst

2

Intra-abdominal collections

2

C-spine fracture

1

3.Limited phase creep

 

    3.1 omitted 1 phase

Abdominal cancer

22

Liver nodule/cyst

2

Hepatobiliary infection

2

Intra-abdominal collections

1

Intra-abdominal foreign body

1

Kidney injury

1

    3.2 omitted 2 phases

Abdominal cancer

5

Liver nodule/cyst

3

Kidney injury

1

Post-operative pain

1

Intra-abdominal collections

1

Adrenal gland nodule

1

Pulmonary tuberculosis

1

    3.3 omitted 3 phases

Abdominal cancer

2

Adrenal gland nodule

2

Intra-abdominal infection

1

150

Table 2. Mean difference and percentage of dose reduction.

n = number of examinations, SD = Standard deviation, * = Statistical significance

Dose lowering technique

(n)

Mean difference (SD)

95% Confidence interval

Dose reduction (%)

p

1.Optimized technical scanner settings

 

   1.1 decreased tube current (81)

526.6 (470.1)

422.7 – 630.6

38.6

<0.001*

   1.2 decreased tube potential (14)

643.8 (404.8)

409.4 - 876.9

47.9

<0.001*

2.Limited image creep (8)

1152.9 (413.1)

807.5 – 1498.3

39.1

<0.001*

3.Limited phase creep

 

 

 

   3.1 omitted 1 phase (29)

1295.9 (721.6)

1021.4 – 1570.4

36.7

<0.001*

   3.2 omitted 2 phases (13)

2767.2 (1437.0)

1898.8 – 3635.5

58.7

<0.001*

   3.3 omitted 3 phases (5)

3103.9 (1779.2)

894.7 – 5313.0

62.1

0.018*

Figure 1. Number of follow-up cases at Burapha University hospital.